2" HEAD DIAMETER - Testing Thermometer with External Adjustment






All 304 stainless steel construction
2" head diameter
Standard stem length: 8", additional stem lengths available by special order
Stem diameter: .156"
Scale length: 4.5"
Rated Accuracy: ± 1% of total scale range
Glass windows are standard, plastic available on special order
Recalibration feature

Model 4200 can be recalibrated by means of an external adjustment by firmly holding the head and immersing the stem at least 2" in an agitated known controlled temperature bath and turning the hex nut at the base of the thermometer head until the pointer indicates the correct temperature.





   Model 4200
Stem Length ==> 8" Spiked Stem length
Range(°) Model / Catalog No.
-100 to 100 °F 4200-08000AF
25 to 125 °F 4200-08000NF
-40 to 160 °F 4200-08000EF
0 to 180 °F 4200-08000HF
0 to 220 °F 4200-08000KF
50 to 250 °F 4200-08000PF
50 to 300 °F 4200-08000QF
50 to 400 °F 4200-08000RF
50 to 500°F 4200-08000SF
   
-100 to 40 °C 4200-08000AC
0 to 50 °C 4200-08000HC
-50 to 100 °C 4200-08000BC
-10 to 110 °C 4200-08000GC
0 to 100 °C 4200-08000IC
0 to 150 °C 4200-08000JC
0 to 250 °C 4200-08000LC

Note: For correct temperature indications, the stem should be immersed at least 2" in fluids and 4" in gases. Head temperature should not exceed 220°F (104°C) .   